OPERATION MANUAL-ECHO-FLEX
™ 5002
© Copyright IBIOM Instruments ltd 2021
Page 8 of 31
Revision: January 2021
6.
OPERATION INSTRUCTIONS
6.1 Procedure after storage
When table is stored for a long period of time, make sure to do this simple procedure
BEFORE
using the table.
1. Plug table to wall outlet.
2. Move down height of table at its lowest position (until it stops by itself) in order to reset
position of electric column. One of next situation may occur:
a. Control box will "
beep
" but table will still move down. Continue to lower table and
it should stop beeping when reaching lowest position. After reaching lowest
position and if it still beeps, hold down 2 buttons together of Trendelenburg
adjustment (3
rd
row), at the same time, until it stops beeping. It will reset all errors
in control box.
b. Control box will "beep" but table will not move. Make sure to check all cable
connection. Hold down 2 buttons together of Trendelenburg adjustment, at the
same time, until it stops beeping. Then, lower down table. If table won't move and
it still beeps, contact us and we will assist you.
c. Same as b, but control box does not make any noise. Follow same procedure.
3. Do an operation test:
a. Move up and down completely height of table;
b. Move up and down completely backrest;
c. Move up and down completely Trendelenburg and reverse Trendelenburg.
4. If table is equipped with a battery (option), leave table plug to the wall for 12 hours to
recharge battery.
5. Table is ready to use.
BATTERY