Instructions

µ–Slide VI

0.4

The ibidi product family is comprised of a variety of µ–Slides and µ–Dishes, which have all
been designed for high–end microscopic analysis of fixed or living cells. The high optical
quality of the material is similar to that of glass, so you can perform all kinds of fluorescence
experiments with uncompromised resolution and choice of wavelength.

The convenient six channel format of the µ–Slide VI

0.4

is ideal for static cell cultivation and

the application of standard immunofluorescence protocols, like treatment, staining, and mi-
croscopy of living or fixed cells. Alternatively, the µ–Slide VI

0.4

can be connected to a pump

and enables you to observe cells under flow conditions.

Material

ibidi µ–Slides, µ–Dishes, and µ–Plates are made of a plas-
tic that has the highest optical quality. The polymer cov-
erslip on the bottom exhibits extremely low birefringence
and autofluorescence, similar to that of glass. Also, it is
not possible to detach the bottom from the upper part.
The µ–Slides, µ–Dishes, and µ–Plates are not autoclavable,
since they are only temperature–stable up to 80°C/175°F.
Please note that gas exchange between the medium and
incubator’s atmosphere occurs partially through the poly-
mer coverslip, which should not be covered.

Optical Properties ibidi Polymer Coverslip

Refractive index n

D

(589 nm)

1.52

Abbe number

56

Thickness

No. 1.5 (180 µm)

Material

polymer coverslip

Please note! The ibidi polymer coverslip is compatible
with certain types of immersion oil only. A list of suit-
able oils can be found on page 3.

Shipping and Storage

The µ–Slides, µ–Dishes and µ–Plates are sterilized and
welded in a gas-permeable packaging. The shelf life under
proper storage conditions (in a dry place, no direct sun-
light) is listed in the following table.

Conditions

Shipping conditions

Ambient

Storage conditions

RT (15-25°C)

Shelf Life of Different Surfaces

ibiTreat, Glass Bottom, ESS

36 months

Collagen, Poly-Lysine

18 months

Fibronectin

4 months

Geometry of the µ–Slide VI

0.4

The µ–Slide VI

0.4

provides a standard slide format accord-

ing to ISO 8037/1. The lateral adapter to adapter distance
of 9 mm (like 96 well plates) allows using multichannel
pipettes.

Dimensions

Number of Channels

6

Channel volume

30 µl

Channel length

17 mm

Channel width

3.8 mm

Channel height

0.4 mm

Adapters

female Luer

Volume per reservoir

60 µl

Growth area

0.6 cm

2

per channel

Coating area using 30 µl

1.2 cm

2

per channel

Bottom matches coverslip

No. 1.5

µ–Slide Surfaces

Depending on the type of cells and the special application
you are using, you will need µ–Slides with different sur-
faces. If you do not require any special adhesion molecules
for your application, the best choice will be ibiTreat, a tis-
sue culture treated surface.

The uncoated µ–Slide is manufactured from hydrophobic
plastic. For the cultivation of most cell lines, it is indis-
pensable to treat the uncoated µ–Slide with biopolymers,
which mediate cell adhesion and growth.

µ–Slide VI

0.4

Page 1

Version 3.0 (2015-12-10)

Summary of Contents for m–Slide VI 0.4 Series

Page 1: ...ymer coverslip Please note The ibidi polymer coverslip is compatible with certain types of immersion oil only A list of suit able oils can be found on page 3 Shipping and Storage The µ Slides µ Dishes and µ Plates are sterilized and welded in a gas permeable packaging The shelf life under proper storage conditions in a dry place no direct sun light is listed in the following table Conditions Shipp...

Page 2: ... confluent layer within 2 3 days Apply 30 µl cell suspension into the channel of the µ Slide Quick dispensing helps to avoid trapped air bubbles Cover reservoirs with the supplied lid Incubate at 37 C and 5 CO2 as usual Await cell attachment in order not to flush out the cells Afterwards fill each reservoir with 60 µl cell free medium Tip The day before seeding the cells we recommend plac ing the ...

Page 3: ...s possible Immersion Oil When using oil immersion objectives use only the im mersion oils specified in the table The use of a non recommended oil could lead to the damage of the plastic material and the objective Company Product Ordering Number Zeiss Immersol 518 F Zeiss 444960 Zeiss Immersol W 2010 Zeiss 444969 Leica Immersion liquid Leica 11513859 µ Slide VI 0 4 Page 3 Version 3 0 2015 12 10 ...

Page 4: ...air cells nerve terminals and ganglion cells Neuroscience 2007 A Lorentzen J Bamber A Sadok I Elson Schwab and C J Marshall An ezrin rich rigid uropod like structure directs movement of amoeboid blebbing cells J Cell Sci 2011 doi 10 1242 jcs 074849 O Mortusewicz W Roth N Li M C Cardoso M Meisterernst and H Leonhardt Recruitment of RNA polymerase II cofactor PC4 to DNA damage sites J Cell Biol 2008...

Reviews: