ibidi Heating System Multiwell Plate – Silver Line
Table 1: (continued)
Thickness of ITO coating
100 nm
Passivation layer
Thickness: 20–25 nm
Refractive index ITO with passiva-
tion layer
1.95
Requirements for Objective Lenses
Minimum 1 mm working distance for coverslip bottom plates (e.g., ibidi µ-Plate 96 Well)
Minimum 2 mm working distance for 0.7 mm thick bottom plates
The sample focal plane in the Incubation Chamber Multiwell Plate lies up to 3.5 mm above
the usual focal plane of the sample. Therefore, the objective must have at least 3.5 mm of
travel distance along the focal axis for imaging.
Microscope Requirements
Working distance condenser
≥
26 mm
Stage opening
160 mm × 110 mm
1.5
Disclaimer
• ibidi shall not be held liable, either directly or indirectly, for any damage incurred as a result of
product use.
• The contents of this manual are subject to change without notice for product improvement.
• This manual is considered complete and accurate at publication.
• This manual does not guarantee the validity of any patent rights or other rights.
• If an ibidi software program doesn’t function properly, this may be caused by a conflict from
another program operating on the computer. In this case, take corrective action by uninstalling
the conflicting product(s).
• ibidi is a registered trademark of ibidi GmbH in Germany and other countries.
1.6
Safety Considerations
WARNING
• Only operate the ibidi Heating System Multiwell Plate with the supplied components.
• Only use the cables and plugs delivered with the system. The power plug of the control unit
must be inserted in an outlet with a ground (earth) contact.
Instruction Manual