ibidi Gas Incubation System – Silver Line
8
Troubleshooting
8.1
Controller Display Errors
Error Description
Possible Cause
Possible Solution
The display does not work, although
the power supply is connected.
Power supply interrupted.
Check the power supply
module and the wire.
ON/OFF switch of gas
mixer off.
Switch on gas mixer.
The humidity channel displays “N.C.”
(Not Connected) in the device display
or IncuControl set up.
Incubation Chamber with
internal humidity sensor
not connected (Figure
Connect
the
Incubation
Chamber as shown in Figure
Humidifying Column channel dis-
plays “N.C.” (Not Connected) in the
device display or IncuControl set up
Humidifying Column not
connected.
Connect
the
Humidifying
Column.
O
2
channel reads “N.C.” (Not Con-
nected) in the IncuControl software or
does not appear on the device display.
Gas mixer for CO
2
only.
Make sure the Gas Mixer is
a CO
2
/O
2
mixer. CO
2
only
mixers do not have an O
2
channel.
The CO
2
or O
2
gas channel displays
large negative values or the error mes-
sage “SHORT” is displayed.
Contact
8.2
Flow Readings
Error Description
Possible Cause
Possible Solution
Gas flow reading is ”0”.
Gas supply lines are not
opened.
Open gas lines.
Gas supply lines are not
connected correctly.
Inspect all gas supply lines connection
points to ensure they are connected securely
and correctly.
Connected gas cylinder is
empty or not opened.
Open or replace the gas cylinder.
Insufficient gas pressure.
Set every gas pressure to 1 bar (14.5 psi).
Device
or
gas
supply
channels are set “STOP”
or “OFF”.
Set “RUN” (Figure
) or “ON” (Figure
using the controller display or IncuControl
software.
Instruction Manual
37
Version 1.0 (June 1, 2022)