Press  the  WAIT  button  to  confirm  all  the  adjusted values:  at  this  time  it  will be  not 
possible to modify the treatment time and the power. Press the START button to start 
the treatment. 

 

When the treatment will be finished, the machine will emit an intermittent signal. If 
you need to stop the treatment before the end press the STOP button and the device 
will stop immediately the laser emission.  

 

In case of an emergency stop press the red button placed on the back of the machine. 

 

After the treatment switch off the device moving the key switch on the OFF  position 
(upper picture).  

Take away the key from the machine to avoid the use by unauthorized personnel.   

Please  do  not  switch  off  the  device  unplugging  the  cable  from  the  power  socket  to 
avoid any type of damage to the device. 

 

Therapeutic indications

 

The following table  shows  the suggested  values for the main therapeutic 
applications considering 20 J/cm

2

 and with a continuous emission)   

The same values can be applied with a pulsed emission (by changing the duty 
cycle) with longer treatment time.   
The given indications can be changed depending of the discretion of the therapist. 

 

 
 

PATHOLOGY 

POWER Watt 

TIME for each 

point         (1 

cm

2

TOTAL TIME  

(4 trigger points )  

 
PAIN RELIEF (acute pain) 
 

 

0,1 

 

200 sec 

 

800 sec 

 
PAIN RELIEF (chronic pain) 
 

 

0,5 

 

40 sec 

 

160 sec 

 
NEK PAIN (acute pain) 

 

0,1 

 

200 sec 

 

800 sec 

Summary of Contents for LA500

Page 1: ...I A C E R Srl MNPG127 02 06 06 17 LASER THERAPY...

Page 2: ......

Page 3: ...E 8 INDICATIONS 8 CONTRA INDICATIONS 8 DEVICE DESCRIPTION 9 PATIENTS PREPARATION 9 HOW TO USE IT 9 Therapeutic indications 11 MAINTENANCE 13 HANDLES CLEANING 13 WORKING PROBLEM 13 ELECTROMAGNETIC INTE...

Page 4: ...important information and useful tips for operating the equipment CAUTIONS The CAUTION message appears before operations which if not correctly performed may cause damage to the machine and or its acc...

Page 5: ...t use the equipment in places where it might get wet Before operating the machine carefully check the correctness of the connections according to the instructions Do not use accessories other than the...

Page 6: ...ecution of different procedures from those specified in this user manual can cause the exposition to dangerous radiation The operator has the responsibility to verify that the issuing head remains wel...

Page 7: ...er production of ATP The applications of the I TECH LA500 laser produce a number of effects on the treated tissues 1 increase in hematic flow vasodilatation of capillaries and arteries 2 biostimulatio...

Page 8: ...ensitive to laser radiation and can be permanently damaged from direct or reflected laser beams The special safety glasses must be worn by both patient by the operator pregnancy the laser is contraind...

Page 9: ...of the patient for the therapy carry out the following operations 1 wash the skin using soap and mild soap or alcohol where the head of the laser will be positioned 2 dry well the skin HOW TO USE IT...

Page 10: ...ent time last 3 minutes Select the operation mode CONTINUOS or CYCLE using the buttons on the right side of the front panel picture below With CYCLE mode it is possible to select the ON time from 1 to...

Page 11: ...ition upper picture Take away the key from the machine to avoid the use by unauthorized personnel Please do not switch off the device unplugging the cable from the power socket to avoid any type of da...

Page 12: ...ITIS 0 3 66 sec 264 sec BAKER S CYST 0 3 66 sec 264 sec DE QUERVAIN SYNDROME 0 2 100 sec 400 sec OEDEMA 0 3 66 sec 264 sec HEMATOMA 0 3 66 sec 264 sec EPICONDYLITIS 0 4 50 sec 200 sec HERNIATED DISC 0...

Page 13: ...ent HANDLES CLEANING The laser handle is very weak and it needs a daily cleaning The following advices are absolutely important to not damage the lens and the laser probe It is suggested to 1 Remove t...

Page 14: ...that the socket works properly The power cable is not inserted correctly into the connector of the device Insert correctly the plug and the cable in the connector of the device The power cable is wor...

Page 15: ...B Output channels 1 Class of isolation parts applied according to the rule EN 60601 1 I BF Protection level against the liquids penetrations according to the rule EN 60601 1 IPX0 Command of execution...

Page 16: ...that the product shall not be disposed of as domestic waste The user must dispose of scrap equipment by taking it to a recognized centre of recycling of electrical and electronic equipment Appendix B...

Page 17: ...ctrostatic discharge ESD IEC 61000 4 2 6kV contact 6kV contact Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30...

Page 18: ...th accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the ME EQ...

Page 19: ......

Page 20: ......

Page 21: ...nd warehouse 30030 MARTELLAGO VE Via S Pertini 24 A Tel 0039 041 5401356 Fax 0039 041 5402684 Cod Fisc P IVA IT 00185480274 R E A VE N 120250 M VE001767 e mail iacer iacer it Internet www itechmedical...

Reviews: