i-OnTemp User Manual
Rev 1.0
Pg. 7
3.
Daily Operations
3.1 i-OnTemp initialization
•
i-OnTemp touchscreen device can be turned on from two different modes:
o
Turn on from total shutdown: Press the on bottom for 5-7 seconds and
wait for the system to boot
o
Turn on from Sleep mode: press the on button for 1 second
•
From the lock screen, swipe up with your finger from the bottom of the screen
towards the top
•
Press the ThermApp icon for camera initialization
•
Wait 15 minutes for the camera to warm up
3.2 Define the Alert temperature
-
Choose a “healthy person” and make him stand at the verification area
-
Check skin temperature and consider it as your reference temperature
-
Update the alarm temperature: For example, 1.5º Celsius or 3º Fahrenheit more
than the reference temperature
The system is ready!
3.3 Fever confirmation
•
If a high temperature is recognized, the system will alert you.
•
As skin temperature can be influenced by environment and personal activity it is recommended to confirm the high temperature with a
manual thermometer before blocking a person’s entrance
3.4 System shut down
•
If the system is used in a fixed place, you do not need to turn the system off – you can leave it plugged in and let it run with no
interruptions
•
While moving the system around, unplug the system and shut it down as follows:
o
Total Shut down: Press the on button for 5-7 seconds and confirm shut down
o
Sleep mode: press the on button for 1 second