Configuration
3.1.4
Parameters
§
Global Parameters
: The measuring time intervals are set here.
§
Period
: Enter measuring interval. The minimum for this parameter is 1 minute for GPRS-transmission. If
Monitoring on the site, a 2 second interval can be used in combination with the live mode (i.e. pumping tests) If the
interval is set as min, sec or hour can be selected from the drop-down menu next to the field. If you set an interval
of less than 1 minute, the program will refuse to transfer the parameters to the logger if the transmission intervals
are too large to all data measured in the interval to be sent in a single transmission.
§
Period-Offset
: From a period of 5 minutes or more, the logger automatically selects the first measurement, the
next "modular" time, i.e. at a period of 1 h, this would be the next full hour. If for example a period of 12 hours is
selected, the measurements would be made, at 0:00 and 12:00. With the "Period Offset" the time of measurement
can be postponed. If an offset of 2 hours is selected for example, the measurements would be made at 2:00 am
and 2:00 pm.
§
Alarm period
: Interval of measurements in case of an alarm
§
Name
: Custom name for each logger (for example, tag number, street name).
IMPORTANT
: The name cannot
contain special characters.
§
Live mode (Real-Time)
: The live mode is used for continuous local transmission of data via radio to a receiver.
When the real time mode is active, the device sends the currently measured data to the shell software which
displays this data in the main window. This allows to monitor the water level in real-time, for example when
changing water levels by pumping. The shortest possible measuring period is two seconds.
§
There are 3 modes for the live mode: