Hypersen Technologies Co., Ltd.
HPS-FT-USER V1.8
30
/
38
3. Select the appropriate port.
4. If the sensor with an RS485 interface is purchased, the default baud rate is 115200. If the
sensor with an EtherCat or EtherNet adapter is purchased, the default baud rate of the sensor is
1500000.
5. Click
OK
to enter the main interface of the host software. Click the
Connect button
to connect
the sensor.
6. If it is prompted that the connection fails, it means that the power supply, wiring, port or baud
rate of the sensor is incorrect. Please check it before reopening the host software for connection.
5.2.2 Filter
Use a low pass filter inside the sensor.
Steps: 1. Complete the connection according to 5.2.1.
2. Click the
Stop
button to stop acquiring data.
3. On the
Parameter
bar of the main interface, move the
Filter
slider to set the filter level.
4. Click the
Start
button to start acquiring data.
5. Check the filtering effect and the change of the communication rate in the “Plot” interface.