background image

www.pacificlab.com.au 

Free Call: 1800 723 405  

   Ph: (03) 9845 0300   

   Fax: + 61 3 9845 0350

  

 

   e-mail:

 

[email protected]

Distributed by

 

 

 

INS0063 

FEB  2015 

Rev. I 

Calibration Control Kit 

Calibration verification of Hygiena SystemSURE Plus and EnSURE luminometers 

Part No# PCD4000 (Kit Includes: 1 Positive Rod and 1 Negative Rod) 

Description/ Intended Use: 

Calibration control rods are used to verify calibration of Hygiena luminometers. 

Positive Rod consists of C

14

 radioactive source that emits very low level of 

low-energy 

β

 radiation in a plastic scintillation matrix. Matrix is configured to 

give constant light output at a high level for up to five years. Negative Rod is 

used to check for possible background light getting into instrument or light 

detector not calibrating correctly. Negative Rod produces low-light (0

4) RLU 

background. It is recommended to verify calibration with Calibration Control Kit 

each week or every other week depending on use and Quality Control 

Program requirements. Incorporating Calibration Control Kit into overall 

Quality Control program will validate luminometer is within specifications and 

operating correctly. 

Directions: 

Positive and Negative Rods are inserted and read in luminometer in same 

manner as ATP test devices. 

Calibration rods do not need to be activated 

like ATP test devices. Do not activate Calibration rods.

 

1.  Open sample chamber and insert Positive Rod. Push rod in gently so it 

makes contact with bottom of chamber.  

2.  Close lid and p

ress “OK” to initiate measurement.

 

3.  Record result. Repeat steps 1 and 2 two more times and record results. 

Calculate average of three readings and enter in table provided below.  

Record Positive Rod Reference No. Refer to Interpretation of Results. 

4.  Repeat steps 1 through 3 using Negative Rod. 

5.  Repeat steps 1 through 4 for each luminometer. Record readings in table 

provided below. Additional tables are printed on back side of this sheet. 

 

Instrument Serial No.: 
Positive Rod Reference No.: 

Test 1 

Test 2 

Test 3 

Average 

 

 

 

 

Negative Rod 

Test 1 

Test 2 

Test 3 

Average 

 

 

 

 

Interpretation of Results: 

Refer to Table 1 below for acceptable RLU ranges. Results should not vary by 

more than ± 20% from Positive Rod Reference No. If results from Positive Rod 

are outside RLU range, or vary by > ±20% from Positive Rod Reference No., 

contact your Hygiena representative. 

 

 

 

 

 

Table 1: In-Calibration RLU Ranges for Hygiena Luminometers  

 

EnSURE (RLU) 

SystemSURE Plus (RLU) 

Negative Rod 

0-4 

0-4 

Positive Rod 

80-160 

40-80 

Incorporating Calibration Control Kit into a Quality Control Program: 

Calibration Control Kit is designed to be incorporated into a Quality Control 

Program that monitors and tracks performance of Hygiena luminometer and/or 

devices. It is recommended that a specific program point be assigned in 

luminometer for results. All results can be viewed directly from luminometer or 

in SureTrend data analysis software. 

Troubleshooting:  

Several factors can cause RLU readings to be outside acceptable range. 

Possible causes include: 

 

Dirty calibration rods. 

Clean outside of both rods with lint-free cloth that 

has been dipped in isopropyl or ethyl alcohol. Air dry and repeat reading.  

 

Damaged calibration rods.

 Return damaged calibration rods to 

Hygiena for replacement. 

 

Luminometer contamination. 

Results that have changed suddenly, 

especially immediately after a number of sampling devices have been 

run, can indicate contamination problems. Results from calibration rods 

can increase or decrease as a result of foreign materials disturbing 

optics chamber of luminometer. If contamination is suspected in 

luminometer, remove read chamber and clean. Scratches on bottom of 

read chamber do not affect readings. Refer to 

owner’s manual for 

instructions on cleaning read chamber. 

 

Position of rod. 

It is important to make sure calibration rods are placed 

in read chamber so rod touches bottom of chamber.  If rod does not 

reach bottom of chamber, rod will be too far from sensor, and will result 

in inaccurate results (low measurements).

 

Storage & Shelf Life:

  

 

Five (5) year shelf life. Refer to expiration date printed on label. 

 

Store at 15 

 25 °C (59 

 77 °F) in box rods were received in, or dark 

container. Store away from light. 

Precautions: 

 

Limit exposure to light. 

Store calibration rods in dark or weak ambient 

light. Kit box provided is recommended as appropriate dark storage. Do 

not expose to direct sunlight or intense artificial light for too long before 

use. 

 

Handle with care. 

Avoid crushing, cutting, or dropping calibration rods. 

Any damage that changes shape or light transmission properties of 

Positive Rod can affect assigned values. Do not use if scintillator tip of 

device is cracked or damaged. 

 

Radioactive material.

 Not for human use. Introduction into foods, 

beverages, cosmetics, drugs, or medicinal products, or into products 

manufactured for commercial distribution is prohibited. Exempt quantities 

should not be combined. 

 

For further safety instruction, refer to Calibration Control Kit Safety Data 

Sheet (SDS). 

 

Disposal: 

 

U.S. & Canada:

 Level of radioactive material in Positive Rod does not 

require special hazardous waste treatment, per U.S. Federal regulations 

governing use and transfer of exempted radioactive materials. Contents 

of Positive Rod are exempt from NRC or Agreement State licensing 

requirements. Positive Rod contains extremely small amounts of 

radioactive Carbon-14, completely contained by outer housing of device. 

There are no radioactive emissions on surface of device, and no special 

precautions are needed in handling device to avoid exposure.  

 

To dispose of Positive Rod, remove or deface “Caution Radioactive

 

Materials” label on rod. 

Positive Rod may then be disposed in trash.  

 

For disposal requirements in other countries, check with your national or 

local supplier.  

 

 

Reviews: