Configuration and Calibration Guide HD0679 Rev 1.2.0 25
Moisture
percentage value
Sensor Unscaled Value
5 Unscaled units = 1.7%
5 Unscaled units = 0.4%
B
A
Figure 10: Relationship of Unscaled Values to Moisture
To determine which mode is the most appropriate it is recommend to run trials for a given
material, mixer type or application. Before doing so it is recommend that you contact Hydronix
to seek advice on our recommended settings for your given application.
Trials differ dependent upon the application. For a measurement taken over time it is
recommended to record the sensor’s output from each of the different measurement modes in
the same process. Data can easily be recorded using a PC and the Hydronix Hydro-Com
software; these results may then be plotted to ascertain the most suitable measurement
mode.
For further analysis, including sensor filtering analysis Hydronix can also offer advice as well
as software to enable the experienced user to achieve the best possible settings for a sensor.
Hydro-Com software and the user guide may be downloaded from www.hydronix.com.
When using the sensor to obtain an output signal that is calibrated to moisture (an absolute
moisture measurement) it is recommended to calibrate using the different measurement
modes and to compare results (see Chapter 3 for more details).
For further information please contact the Hydronix support team at [email protected]
Summary of Contents for Hydro-Mix
Page 4: ...4 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 6: ...6 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 8: ...8 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 10: ...10 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 14: ...Chapter 1 Introduction 14 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 28: ...Chapter 2 Configuration 28 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 42: ...Chapter 4 Best Practices 42 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 48: ...Chapter 5 Sensor Diagnostics 48 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...
Page 52: ...Appendix A Document Cross Reference 52 Configuration and Calibration Guide HD0679 Rev 1 2 0 ...