8.5
Electrical power supply of Display A
Risk of electrocution. Always disconnect the power supply before making any electrical
connections.
Potential damage to the device. The internal electronic components of the device can
be damaged by static electricity, which might affect its performance characteristics and
operation.
Display A can be powered according to one of the following five modes:
Internal 3.6 VDC lithium battery pack
External 3.6 VDC lithium battery pack
Internal 9 to 28 VDC alkaline battery pack
An external 9 to 28 VDC power supply connected to the labelled 9-28 VDC terminals
A 9 to 28 VDC external power supply connected to the military connector
These different power supply modes can be implemented using the two elements illustrated below:
Switch SW2 with two positions:
Internal
or
External
Connector J15