Configuring the SMU
SensorMonitoring Unit - SMU
en(us)
Page 60 / 100
BeWa SMU1200 3854281 V2.0x en(us) 2013-08-20
2013-08-20
SEN.ADR – Set sensor address
Use this menu item to reset the sensor address of the connected sensors.
This becomes necessary if an AS or another sensor at the sensor interface B
is used without a fixed address or with the same address as at sensor
interface A.
To change the sensor address, proceed as follows:
Connect the CS1000 or MCS to
sensor interface A and the
AquaSensor AS to sensor
interface B.
Call up the PowerUp menu.
Change to menu item
o.k.
Request changes to
the address settings
Esc
Cancel and back
Contamination
SAE/NAS
SEnADR
The SMU determines the address of
the sensor connected to sensor
interface A.
Remove the sensor from sensor
interface A and confirm with OK.
o.k.
Confirm
Esc
Cancel and back
Contamination
SAE/NAS
REMOVE
SEnA
ok
The settings for the sensor at
sensor interface B (AS1000) will
now be reset.
The message WAIT appears on the
display.
Contamination
SAE/NAS
WAIT
The bus address of the sensor at
interface B is set automatically.
The AS is set as bus address B. If
this bus address has already been
allocated, bus address C is selected
for the AS.