Measuring menu
MCS1xxx /-T005
en(us)
Page 51 /
68
BeWa MCS1xxx-T005 4668567a en-us 2022-06-14.docx
2022-06-14
STATUS: Status byte
(00 at status = O.K.)
FI:
Field strength of the field coil in the MCS
TEMP C: Medium temperature in °C (indirect
measurement via sensor in the spool)
TEMP F: Medium temperature in °F (indirect
measurement via sensor in the spool)
Factory setting:
FE A
The value of 1000 µm is shown on the display as 1k.
SWT.OUT – Set the switching output
Function:
You can find a detailed description on page 38.
Parameter:
FE.NFE S1 = ferromagnetic (Fe)
S2 = Non-ferromagnetic (nFe)
ALL.RDY S1 = ferromagnetic (Fe) and non-
ferromagnetic (nFe)
S2 = Device Ready
ALARM
S1 = Alarm
S2 = Device Ready
Factory setting:
FE.NFE
SWT.LOG - Set logic at the switching output
Parameter:
ACT.HI
active high
ACT.LOW active low
Factory setting:
ACT.HI
SWT.PLS - Set pulse length at the switching output
Function:
Setting the pulse length of the switching outputs
Parameter:
5 - 200 milliseconds
Factory setting:
7
SWT.TST – Set test pulse at the switching output
Function:
Function test of the switching outputs