FluidAqua Mobil - FAM 10
Troubleshooting
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 45 / 72
BEWA FAM10 3102456f en-us 2012-03-13.doc
2012-03-13
Level switch 01/02(02/03)
Step
Description
Malfunction:
Level switch 01/02(02/03)
1.
Switch off the FAM at the main switch.
2
Remove the connector to the level switch and then remove the level
switch.
3.
Manually check the function of the switching points.
->a. Switching points are available
-> proceed to Step 5.
4.
->b. Switching points not available
-> proceed to Step 6.
5.
Inspect the plug and the cable to and in
the control cabinet for the PLC.
Functioning?
-> proceed to Step 7.
6.
Replace the level switch.
->a. The function is available.
-> proceed to Step 8.
7.
->b. The function is not available.
-> proceed to Step 9.
8.
The PLC is faulty.
9.
Replace the faulty parts.