6 | Operation
HYDAC FILTER SYSTEMS GMBH
62 / 124
BeWa CTU1x3x-CTU1x4x 4093420f en-us web
NOTICE
Ultrasonic extraction without a suitable device
The the analysis chamber will be damaged
u
Fill adequate test liquid in the analysis chamber.
u
Place the test item in the analysis chamber only with the corre-
sponding jig.
u
Use the hook-in basket. Note the permissible weight of the test
item of 0.5 kg.
For the ultrasonic test, replace the grating in the analysis chamber
with a proper device, such as a freely hanging basket.
Hang the basket retaining frame (a) with the basket (b) in the cor-
responding thrust bolts in the analysis chamber.
b
a
Fig. 27:
Analysis chamber with mounts
(a) Thrust bolts with basket retaining frame
(b) Basket for ultrasonic extraction
1
2
3
4
5
6
7
8