ContaminationSensor CS 1000
Analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 59 / 108
BeWa CS1000 3247149p 240 en-us 2011-02-11.doc
2011-02-11
SAE + T
The SAE+T signal consists of 5 measured values which are transmitted time-coded
with the following time slices:
4,0
2,0
I (mA)
1
2
4
6
8
10
3
5
7
9
U (V)
time (ms)
4,8
19,2
High
High
Low
Low
300
3000
3000
3000
3000
2,25
2,4
9,6
4,5
9,75
19,5
9,9
10,0
9,85
19,7
0,0
19,8
_
Ti
me
Signal
Size
Signal
duration
per pulse
in ms
Current (I) / Voltage (U)
1
Identifier
SAE A
300
High / Low
2 Measured
value
SAE A
3000
Current/Voltage for measured value
3
Identifier
SAE B
300
High / Low / High / Low
4 Measured
value
SAE B
3000
Current/Voltage for measured value
5
Identifier
SAE C
300
High / Low / High / Low / High / Low
6 Measured
value
SAE C
3000
Current/Voltage for measured value
7
Identifier
SAE D
300
High / Low / High / Low / High / Low / High / Low
8 Measured
value
SAE D
3000
Current/Voltage for measured value
9
Identifier
Temperature
300
High / Low / High / Low / High / Low / High / Low /
High / Low
10 Measured
value
Temperature 3000
Current/Voltage for measured value