Technical data
Technical data
Measured value display
Continuous view in LCD display
Self-diagnosis
Continuous self-monitoring with error
indication in LCD display
Particle size channels
FCU 20xx-x:
5µm / 15µm / 25µm / 50µm
FCU 21xx-x
:
2µm / 5µm / 15µm / 25 µm
FCU 22xx-x
:
4µm
(c)
/ 6µm
(c)
/ 14µm
(c)
/ 21 µm
(c)
Measurement range (calibrated)
FCU 20xx-x
NAS 2 - 12
ISO 10 / 09 - 21 / 18
FCU 21xx-x
:
NAS 2 - 12
ISO 12 / 10 / 09 - 23 / 21 / 18
FCU 22xx-x
:
SAE 2 - 12
ISO 12 / 10 / 09 - 23 / 21 / 18
Indication range
FCU 20xx-x
NAS 2 - 15
ISO 10 / 07 - 23 / 21
FCU 21xx-x
:
NAS 2 - 15
ISO 12 / 10 / 07 - 25 / 23 / 21
FCU 22xx-x
:
SAE 2 - 15
ISO 12 / 10 / 07 - 25 / 23 / 21
Calibration accuracy
± 1/2 class (ISO,NAS,SAE)
Calibration
FCU 20xx-x
:
ISO 4402
FCU 21xx-x
:
ISO 4402
FCU 22xx-x
:
ISO 11943
Recalibration / Service
Recommended: every 2 to 3 years
Battery-buffered log memory
Can accommodate up to 3000 measured
values.
FCU 2000-4 series
en(us)
99 / 120
BEWA FCU2000-4 3109987h en-us 2017-01-24.docx
2017-01-24