4.6 Technical Data
The FCU has the following technical data.
Contamination sensor
Self diagnostics
continuously with error indication
via status LED and display
Display
LED, 6 digits, with 17 segments
Measured values for
solid particle contami-
nation
ISO code / SAE class
Measurement range
Display of ISO code between:
9/8/7 … 25/24/23.
Calibrated within the range of ISO
code:
13/11/10 … 23/21/18
Accuracy
Contamination sensor: ± 1 ISO code
Service Display
FLOW / DRIVE / Temp
Measuring time
30 seconds
Tab. 4:
Contamination sensor
Hydraulic data
Suitable Fluids
Mineral oil or raffinate based on min-
eral oil
Hydraulic connections
INLET
Test connector type 1604
OUTLET
DN7 nipple socket
Maximum permissible
operating pressure at
INLET connection
-0.5 … 45 bar / 0 … 650 psi
Maximum permissible
operating pressure at
OUTLET connection
0 … 0.5 bar / 0 … 7.5 psi
Measurement flow rate 30 - 500 ml/min (viscosity depen-
dent)