14
will cease to step up and into the state. If no breakdown for 15 s,
the instrument will default to the current stop up voltage of
insulating oil breakdown voltage;
On the default value is 6, optional range 1 ~ 6 times (incremental Δ
= 1);Set after press ok button to return to the start page, press the
start key test;
Cup a cup press this button to enter a selection child pages,
specific see six operation, operation steps diagram 3. The related
content.
7. For this model, every cup of up to 6 times parallel determination
parameters such as the breakdown voltage value will be
automatically stored. After the measurement screen will display to
remind the test, press the confirmation key to return to the start
page (figure 1).Press to print or display, enter the sample single
breakdown voltage value, arithmetic mean value and measurement
of the date and time display child pages (figure 7 ~ 9).
Figure 7.