46
6.4.2.
Capturing anterior segment
1.
The first procedure is same as ‘6.1 General Operation: procedure 2 ~ 7’.
•
On the anterior mode, function of auto shooting and auto tracking is not available.
2.
Move the body to align patient’s eye. Move the body slowly while watching patient’s eye and body,
because working distance is just 15mm that the front lens is very close to patient's eye.
3.
Capture image and check image quality (in anterior radial or anterior 3D mode)
3-1. Alignment and focus
①
Move body with joystick slowly to align anterior scan line (A) and center of patient’s eye
(B)
②
Start OCT scan by pressing scan start icon (
).
③
Move body with joystick slowly until section of cornea appears on the screen.
3-2. Measuring curvature of cornea
①
Optimize OCT signal by pressing optimize icon on the screen (
) or optimize
button on the body (A).