66
Excelon EZ Operator’s Manual
7. System Maintenance
7.1 Frame Reader & Blocker Calibration
Execute Automatic Calibration for Frame Reader & Blocker in the followings.
•
The traced data seems to show the error in the Size and Axis
•
The Frame Reader seems not to work properly
•
Blocking position seems to incorrect
Frame Calibration
Procedures for frame calibration
①
Place standard frame to tracer.
②
Enter menu mode by pressing the
[MENU] button for 2 seconds.
③
Select “User Mode”.
④
Select “Calibration”.
⑤
Select “Frame Calibration”.
⑥
Then follow instructions displayed on the
LCD.
Standard Frame
!
NOTE
•
The result of Automatic Calibration will be automatically memorized.
•
Be sure to check the related information before executing the calibration work
because you cannot get back the machine to the previous status. .
•
During the calibration, [STOP] button stops and aborts processing job.