Sonicaid FM800 Encore Series Instructions for Use
38
5.6
58B
Serial interface
When connecting to a central review system, the appropriate serial interface protocol
must be set to match the software system. The ‘RS232 HP 50’ protocol is the most widely
used and is recommended for use with Sonicaid Centrale, Obstetric Archive and a range
of other CRS systems. For Sonicaid Fetalcare, select the ‘RS232 Sonicaid’ protocol. For
further information, refer to the FM800E service manual.
To select the appropriate serial interface:
1 > HOST INTERFACE.
2 Use the labelled key to select the host interface type.
For Sonicaid FetalCare make sure the setting is ‘RS232 Sonicaid’.
For Sonicaid Centrale, Obstetric Archive and other CRS systems that use the ‘HP 50’
protocol, make sure the setting is either ‘RS232 HP 50’ when connecting via the serial
port. Alternatively, for Sonicaid Centrale, select ‘Network HP 50’ when connecting via
the LAN interface.
3 > EXIT.
4 Switch off FM800E. Wait for about 10 seconds.
5 Switch FM800E on again.
See also Chapter 14.
Note: Sonicaid Centrale can also support the ‘Sonicaid’ protocol option
Summary of Contents for Sonicaid FM800 Series