background image

 

 

 

 
FM800Encore Service Manual  

 

 

 

 

    

 

FM800Encore 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

FM800ENCORE

®

 SERVICE MANUAL  

 
 
 

Summary of Contents for FM800Encore

Page 1: ...FM800Encore Service Manual 1 FM800Encore FM800ENCORE SERVICE MANUAL...

Page 2: ...Mounting 13 6 Safety Testing 16 6 1 Protective Earth Continuity 16 6 2 Earth Leakage Current 17 6 3 Patient Leakage Current 17 6 4 Patient Leakage Current Mains on applied parts 18 6 5 Patient Auxilia...

Page 3: ...cts Although every care has been taken to ensure that the information in this manual is accurate continuous development may result in equipment changes The Company reserves the right to make such chan...

Page 4: ...NING Explosion hazard Do not use FM800Encore in the presence of flammable anesthetics or gases WARNING Do not spray pour or spill any liquid on FM800Encore its connectors switches or openings in the c...

Page 5: ...ch may cause interference or loss of signal A frequent electrical and visual inspection on cables and wires is advisable CAUTION Observe ESD electrostatic discharge precautions when working within the...

Page 6: ...ponents that need to be replaced This document is not intended as a full diagnostic tool because many of the boards and modules can not be satisfactorily repaired in the field and must be returned to...

Page 7: ...core Designed to help with the management and care of all pregnancies where comprehensive and adaptable fetal monitoring facilities are needed the Sonicaid FM820Encore offers twin ultrasound fetal and...

Page 8: ...680 to 1060 mB Storage transport pressure 68 to 106 kPa 680 to 1060 mB Operating humidity 10 to 75 RH non condensing Storage transport humidity 10 to 90 RH non condensing 4 3 AC supply voltage and fus...

Page 9: ...V peak to peak referred to input Safety Type CF protection 4 6 Uterine activity external Toco Range 0 100 relative units Sensitivity 100 FSD equivalent to 120g Offset range 100g Auto zero Manual and a...

Page 10: ...Inflation max 300 mmHg Measurement time max 160 secs Standards ANSI AAMI SP10 1992 4 10 Maternal oximetry Sensor types a The following probe is suitable for FM830Encore equipped with BCI B1 MSpO2 modu...

Page 11: ...ree standing on any suitable stable surface The unit should not overhang the surface There are four high adhesion rubber feet attached to the base of the unit these should help prevent the unit moving...

Page 12: ...Installation FM800Encore Service Manual 12...

Page 13: ...es with comprehensive manufacturers instructions on securely fitting and using the bracket these should be read and followed When the approved wall mounting kit is correctly assembled and attached to...

Page 14: ...Installation FM800Encore Service Manual 14 STEP 2 Adjust the pivot arm to the required height STEP 3 Slide the unit onto the carrier plate Note Make sure the locking pin is fully engaged...

Page 15: ...Installation FM800Encore Service Manual 15 Item Description Quantity 1 MOUNTING PLATE 1 2 M3x14 SCREW 4 3 GCX M SERIES 12 PIVOT ARM 1 4 GCX UTILITY BASKET 1 Locking pin must be fully engaged...

Page 16: ...insulation 6 1 Protective Earth Continuity This test checks the integrity of the power cord ground wire from the AC plug to the instrument chassis ground A current of 1A from a current source with a...

Page 17: ...ine Voltage OS Open Supply Note All measurements shall be made with the power switch in ON position 6 3 Patient Leakage Current This test measures patient leakage in accordance with IEC60601 1 clause...

Page 18: ...voltage will be present on the applied part terminals during this test Exercise caution to avoid electrical shock hazard Do not touch the patient lead clips or the simulator parts connected to patien...

Page 19: ...trical safety analyser as recommended by the analyser operating instructions 2 Connect the monitor s AC plug to the electrical safety analyser as recommended by the analyser operating instructions 3 T...

Page 20: ...onnect the monitor to AC power source using proper power cord 2 Press the ON switch 3 Verify Green power LED is lit 4 Verify Amber printing LED is flashing for approx 2 seconds 7 3 2 Display 1 Check t...

Page 21: ...that after maximum 13 seconds the SpO2 will flash and the alarm will sound The digital readout must indicate no reading and the HEART SYMBOL should flash 5 Attach the finger probe to a QA 510 simulato...

Page 22: ...the back of the hand approximately twice per second see figure below 3 The display should indicate a rate of around 120 BPM after a few seconds Continue this for one minute and check for a correspondi...

Page 23: ...Diagrams FM800Encore Service Manual 23 8 Diagrams Drawing No Description Figure 1 FM800E Block Diagram...

Page 24: ...ment address decoding generation of an interrupt request for the keyboard decoding of the address for the RTC synchronous serial interfaces and clock generation Shared RAM This is to provide a means b...

Page 25: ...T END This is to provide signal amplification and filtering using a high input impedance high common mode rejection amplifier Also provides impedance checking circuit which has been designed to operat...

Page 26: ...26 9 Disassembly Removing the chassis cover The chassis cover is held by a total of 11 screws 7 at the rear panel and 4 on the side rails Torque 60cN M Gently slide the lid to the back and lift upward...

Page 27: ...Service Manual 27 Removing NIBP Module Detach the air tube from the module as shown below Remove the four mounting screws Detach the NIBP loom from the module PCB M3 Screws and Shake proof Washers NIB...

Page 28: ...Disassembly FM800Encore Service Manual 28 Module PCB The module PCB is held by 8 M3 screws as shown below BCI MSPO2 Module Top view Rear view TOP VIEW...

Page 29: ...SPB FM820E or FM830E FM830E side pod FM820E Pod bracket Plastic Cover Strip Ultrasound 1 CON7 Ultrasound 2 CON8 FECG MECG CON9 TOCO CON11 Paper out sensor Print head interface FM830E Front panel FM820...

Page 30: ...ervice Manual 30 Digital board NDB Push the PCB this way 8 off Pillars M3 Screw Paper out sensor Print head interface Remove tie bar screw Unplug event marker connectors Unplug power connectors Unplug...

Page 31: ...py water and dry Patient ECG leads The patient leads should be cleaned with warm water or a neutral cleaner and wiped dry To disinfect use chemical disinfectants containing ethanol 70 80 propanol 70 8...

Page 32: ...Reconnect LK1 and assemble the init Set the correct time and date values CAUTION Discarded battery may explode during incineration Recycle used batteries properly Do not dispose of batteries in refuse...

Page 33: ...transducer placement Is the transducer broken Consider switching from Ultrasound to FECG No FHR trace printed Has the Print button been pressed Is there paper in the paper tray Is the paper tray full...

Page 34: ...unit type serial number date of purchase if known and written details of the symptoms or fault found Orders for spare parts may be sent by post telex fax or telephoned to Huntleigh Healthcare or appro...

Page 35: ...aste electrical and electronic equipment WEEE and that this unit was placed on the market after 12 August 2005 This directive covers EOL end of life disposal Rules for the User within the EU at end of...

Page 36: ...HHC Defective consumables will be replaced Any problems should be reported immediately to the supplier of the equipment or your local HHC service department The warranty is valid in all circumstances...

Reviews: