WARNINGS:
DANGER OF ELECTRIC SHOCK. DO NOT TOUCH ANY EXPOSED PARTS
WHILE CONDUCTING THIS TEST. DO NOT TOUCH ANY PART OF YOUR
BODY WITH THE CONTACTS OF THE PROBES.
THIS EQUIPMENT SHOULD NOT BE USED IF YOU HAVE A HEARING AID
OR PACEMAKER FITTED, DUE TO THE POSSIBILITY OF
ELECTROMAGNETIC DISTURBANCE.
Specification
: 3.0 kVac is applied between the connected Live and Neutral wires and the appliance’s
metal compressor assembly box. No breakdown should occur.
Note : The voltage levels used for this test, may stress and weaken the insulation. This test, is
therefore not recommended as a routine test. It should only be used after a major
assembly/disassembly has been completed.
3
PCB Functional Tests (Electrical Units Only)
Tests must be carried out using a 220-240 V, 50 Hz power supply.
Test Setup and Initial Switch On
3.1
Install the mating parts of the snap-lock connectors on the pump outlets so that both outlets are
open to atmosphere. Set the pressure control knob fully counterclockwise.
3.2
Connect the mains/power cable to a suitable power socket outlet and then set the run/standby
switch to ‘run’. All four indicator lights should illuminate and the audible alarm should sound for
about 3 seconds. The green
ON
indicator, the amber
WAIT
indicator and the
LOW
PRESSURE
indicator will remain on. The
POWER FAIL
indicator light should extinguish.
Power Fail Alarm Check
3.3
With the pump operating, disconnect the power supply at the power socket outlet then
reconnect it after 4 seconds. Verify that no alarms are given and normal operation is resumed.
3.4
Again disconnect the power supply at the power socket outlet. Verify that, after 5 seconds, the
red
POWER FAIL
indicator flashes in conjunction with the audible alarm. The
WAIT
indicator
should go off when the
POWER FAIL
alarm indicator flashes. Reconnect the power supply.
Verify that the
POWER FAIL
indicator is extinguished.
3.5
Operate the run/standby switch to ‘standby’ and verify that all indicators are off. After 5
seconds, operate the switch back to ‘run’. Verify that all alarm indications are cleared and
normal operation is resumed. Operate the run/standby switch to ‘standby’.
Low Pressure Alarm Check
3.6
Use suitable sealing caps to close the ports. Operate the run/standby switch to ‘run’. Verify
that, after 1 minute, the
WAIT
indicator is extinguished.
3.7
Open both ports to atmosphere. Verify that, after 4 minutes and 45 seconds, the red
LOW
PRESSURE
indicator flashes in conjunction with the audible alarm.
3.8
Operate the run/standby switch to ‘standby’ and back to ‘run’. Verify that all alarm indications
are cleared and normal operation is resumed.
Chap 5
Issue 2
Page 6
SER0004
Nov 2001
Testing
Mini Pump Service Manual
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...