33 Replacing the Microswitch
Note : It is not necessary to remove the pressure control module from the pump case.
33.1 Pull off connectors from the Micro switch terminals 1 and 3 (Fig 18, Item 10).
33.2 Release two screws (Fig 18, Item 20) and remove the Micro switch from the pressure control
module.
33.3 Replacement is the reverse of the above procedure.
34 Replacing the Terminal Block
Note : It is not necessary to remove the pressure control module from the pump case.
34.1 Disconnect all wires from the terminal block (Fig 18, Item 30).
34.2 Release two securing screws (Fig 18, Item 40) and remove the terminal block from the
pressure control module.
34.3 Replacement is the reversal of the above procedure. Refer to the wiring diagram (Chap. 4,
Fig 24) when reconnecting the wires for the electronic version of the pump and wiring diagram
(Chap. 4, Fig 25) when reconnecting the wires for the mechanical version of the pump.
Issue 2
Chap 3
Nov 2001
SER0004
Page 35
Mini Pump Service Manual
Pump Repair
Summary of Contents for AlphaBed
Page 1: ...Issue 2 SER0004 Nov 2001 Mini Pump Service Manual Huntleigh HEALTHCARE...
Page 2: ......
Page 4: ...Contents Issue 2 Page ii SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 8: ...Contents Issue 2 Page iv SER0004 Nov 2001 Contents Mini Pump Service Manual...
Page 16: ...Chap 1 Issue 2 Page 8 SER0004 Nov 2001 Introduction Mini Pump Service Manual...
Page 20: ...Chap 2 Issue 2 Page 4 SER0004 Nov 2001 Maintenance Mini Pump Service Manual...
Page 82: ...Chap 5 Issue 2 Page 8 SER0004 Nov 2001 Testing Mini Pump Service Manual...
Page 86: ...Chap 6 Issue 2 Page 4 SER0004 Nov 2001 Technical specification Mini Pump Service Manual...
Page 92: ...Chap 7 Issue 2 Page 6 SER0004 Nov 2001 Parts list Mini Pump Service Manual...
Page 94: ...Chap 8 Issue 2 Page 2 SER0004 Nov 2001 Service Contact Details Mini Pump Service Manual...