UltraScan PRO Supplemental Manual for EasyMatch QC Ver. 2.0
35
UltraScan PRO Specifications
The specifications and characteristics of your instrument are given in this section.
For best performance, your instrument should be placed where there is ample work space with
medium or subdued illumination and no drafts. For optimum results, a clean, air-conditioned area is
recommended with a relative noncondensing humidity of 20-80% [10-85% for sensor only] and
relatively constant temperature not exceeding 32°C (90°F) [38°C (100°F) for sensor only]. For
specification performance, the recommended temperature range is 21-28°C (70-82°F).
The instrument should be connected to a stable, instrument-grade power line. If other equipment is
connected to the same power line, a transient power surge may be produced when the other
equipment is turned on. If this happens, restandardize the instrument before making measurements.
HunterLab recommends using a line conditioner with a minimum 600 VA rating and a battery back-up
system.
Physical Characteristics
Dimensions
Width: 42 cm (16.5 in)
Height: 32.3 cm (12.7 in)
Depth: 49.8 cm (19.6 in)
Weight: 25.9 kg (57 lbs) not including the sample clamp
Environmental Requirements
Operating Temperature
4°C - 38°C (40°F - 100°F)
Operating Humidity
up to 85% relative, non-condensing
Storage Temperature
-21°C - 66°C (-5°F - 150°F) for up to three weeks
Power Required
Voltage
90-250 VAC, 50/60 Hz
Single Phase
100 VA maximum
Fuse
2A, SB (110 V) or 1A, SB (220 V)