SpectraTrend HT User’s Guide 3.2
49
Figure 50. Date/Time
Run Configuration
When
RUN CONFIG
Setting is highlighted press the
RIGHT ARROW
button to move to the
configuration screen. You may set the following parameters by using the
UP
and
DOWN
buttons:
Figure 51. Run Configuration & Read Over Time
RUN CONFIGURATION: READ MODE
You may choose whether to read and present data averaged over a selectable time
period (
TIME
) or averaged for each piece read (
PIECE
) by pressing the
RIGHT ARROW
button.
If
PIECE
is selected, the system waits until the product is sensed, collects data until the
product is no longer sensed, and then transmits
average data
, in units selected, for all
the readings taken on the piece. The maximum speed at which the sensor can measure
in piece mode is 1 per second. This mode is appropriate for individually-measured
products such as cookies or crackers.
If
TIME
is selected, the SpectraTrend HT continuously measures and reports data over
specific periods of time, transmitting
average data
, in units selected, for the selected
time interval. Gaps in product when in time mode are ignored by the sensor. This mode
is appropriate for products measured in piles or batches, such as potato chips or
powders.
In
TIME MODE
, all sets of results for a product are used in the average result for the
time period defined as the
READ TIME
. The default read time is 15 seconds.
Summary of Contents for SpectraTrend HT
Page 6: ...SpectraTrend HT User s Guide 3 2 6...
Page 21: ...SpectraTrend HT User s Guide 3 2 21...
Page 22: ...SpectraTrend HT User s Guide 3 2 22...
Page 23: ...SpectraTrend HT User s Guide 3 2 23...
Page 24: ...SpectraTrend HT User s Guide 3 2 24...
Page 25: ...SpectraTrend HT User s Guide 3 2 25...
Page 26: ...SpectraTrend HT User s Guide 3 2 26...
Page 27: ...SpectraTrend HT User s Guide 3 2 27...
Page 28: ...SpectraTrend HT User s Guide 3 2 28...
Page 76: ...SpectraTrend HT User s Guide 3 2 76...
Page 86: ...SpectraTrend HT User s Guide 3 2 86...