SpectraTrend HT User’s Guide 3.2
32
Note: there is no backspace command. In order to get rid of unwanted spaces
return to the first position in the name field and re-enter the chosen name by
pressing the right arrow button. To delete a character use the right arrow button
to select (highlight) the desired character and then press the go (space) button.
To scroll through the name field/box to go back to the beginning or the desired
position press the right arrow button.
Product Standards
Four types of standards are available:
AD HOC, NUMERIC, HITCH
, and
PHYSICAL
. To select your
standard press the
RIGHT ARROW
key.
Choose
AD HOC
when you wish to use the first measurement value as the standard. This
setting is useful if the standard values are unknown or you need the trending measurement
values from the beginning of the Run. Measurement values from the previous run are
overwritten/lost for each subsequent run using this standard.
Figure 16. Product Setup & Ad Hoc Standard Type
Choose
NUMERIC
when an actual product specimen is not available for measurement, but you
know the target color values taken from previous measurements of the product specimen. The
color values for a numeric standard are entered and stored in the product setup for as long as
desired.
Figure 17. Product Setup & Numeric Standard Type
Summary of Contents for SpectraTrend HT
Page 6: ...SpectraTrend HT User s Guide 3 2 6...
Page 21: ...SpectraTrend HT User s Guide 3 2 21...
Page 22: ...SpectraTrend HT User s Guide 3 2 22...
Page 23: ...SpectraTrend HT User s Guide 3 2 23...
Page 24: ...SpectraTrend HT User s Guide 3 2 24...
Page 25: ...SpectraTrend HT User s Guide 3 2 25...
Page 26: ...SpectraTrend HT User s Guide 3 2 26...
Page 27: ...SpectraTrend HT User s Guide 3 2 27...
Page 28: ...SpectraTrend HT User s Guide 3 2 28...
Page 76: ...SpectraTrend HT User s Guide 3 2 76...
Page 86: ...SpectraTrend HT User s Guide 3 2 86...