EasyMatch QC User’s Manual Version 2.1
Measurement Values
A-31
Coefficient
C/2°
D65/2°
C/10°
D65/10°
C
X
1.2769
1.2985
1.2871
1.3013
C
Z
1.0592
1.1335
1.0781
1.1498
YI E313 Difference is also available, and is defined as follows:
dYI E313 = (YI E313)
sample
- (YI E313)
standard.
Yellowness Index per ASTM Method D1925
is calculated by the software as follows:
(
)
Y
Z
8
1.05839817
-
5X
1.27497679
100
=
D1925
YI
under C/2° conditions for all instruments
except UltraScan XE.
(
)
Y
Z
2
1.05743409
-
6X
1.27464150
100
=
D1925
YI
under C/2° conditions for UltraScan XE.
The yellowness index formula is shown in ASTM D1925 as:
(
)
YI D1925 =
100 1.28 X
- 1.06 Z
Y
CIE
CIE
CIE
under C/2° conditions.
The tristimulus values of clear air (for CIE illuminant C and the 1931 CIE 2° standard observer) are X =
98.041, Y = 100.000, Z = 118.103. Using these values, the ASTM formula yields YI = 0.303 for clear
air because the factors are truncated to three significant figures. In order to set the yellowness index for
air equal to 0.0, the constant multipliers for X
CIE
and Z
CIE
have been expanded slightly.
The ASTM D1925 method was withdrawn in 1995, but this formula still provides useful information.
This index is always calculated for C/2°, regardless what illuminant and observer are chosen.
YI D1925 Difference is also available, and is defined as follows:
dYI D1925 = (YI D1925)
sample
- (YI D1925)
standard.
Paper Brightness (Z%)
Paper brightness, Z%, is used in the evaluation of the degradation of white materials. It can also be a
measure of the effectiveness of bleaching.
Z% =
100Z
Z
CIE
n
.
Z% Difference is also available, and is defined as follows:
dZ% = (Z%)
sample
- (Z%)
standard.
Summary of Contents for EasyMatch QC
Page 11: ...Part I Software Overview...
Page 105: ...EasyMatch QC User s Manual Version 2 1 2 50 Lessons...
Page 112: ...Part II Software Menu Commands...
Page 140: ...EasyMatch QC User s Manual Version 2 1 3 28 The File Menu...
Page 150: ...EasyMatch QC User s Manual Version 2 1 3 38 The File Menu...
Page 156: ...EasyMatch QC User s Manual Version 2 1 5 2 The View Menu Toolbar is not displayed...
Page 227: ...Part III Instruments...
Page 290: ...Part IV Reference...
Page 325: ...EasyMatch QC User s Manual Version 2 1 Measurement Values A 35...