EasyMatch QC User’s Manual Version 2.1
UltraScan VIS
11G-45
10. Plug in the white connector of the new lamp.
11. Replace the cover and screws on the lamp housing.
12. Plug the harness from the instrument into the indicator lights of the front panel.
13. Line the cover up with the guides on the bottom of the instrument.
14. Replace the front cover on the instrument and tighten the two front screws as well as the inner bezel
thumb screw.
15. Turn the power on.
16. Standardize the instrument.
Replacing the Fuses
The fuses are located on the back panel of the instrument. Follow the instructions below to replace the
fuses.
CAUTION
Replace only with specified fuses, HL#13-2600-30, or equivalent. More information
concerning this fuse may be found in the UltraScan VIS Specifications section of this
chapter.
1. Disconnect the UltraScan VIS from its power source by unplugging the power cord.
2. Remove the fuse cartridge from the back panel of the instrument. You may want to use a small, flat
screwdriver to help you pry it out.