HunterLab ColorFlex 13-1 User Manual Download Page 1

07/02

ColorFlex

13-1

ColorFlex

Notice:  Use of this equipment in a manner not specified by the manufacturer may
impair the protection afforded by the equipment.

The ColorFlex system consists of a ColorFlex optical sensor and an IBM-compatible computer
composed of the system unit, monitor, keyboard, and mouse.  The computer may have been user-
provided or purchased with the sensor and software.  A printer may also be used with the system.

The ColorFlex Sensor

The ColorFlex spectrocolorimeter is a versatile, compact color measurement instrument that can be used
on products of virtually any size and in industries as diverse as paint and textiles.  It is available in two
different models, 45°/0° and diffuse/8° (sphere).  The label on the bottom of the instrument identifies the
geometry.  The instrument is designed for use with the sample port facing up or forward.

The Read Button

The 

Read

 (lightning bolt) button on the keypad, also known as the “Sensor button,” can be used to

perform any command or macro assigned to it through the 

Toolbar

 command on the 

Configure

 menu.

By default, this button activates a sample measurement.  To assign it another function, see the Toolbar
section in the Configure chapter, Chapter 9.

LCD screen

(display)

Sample port

Keypad

Summary of Contents for ColorFlex 13-1

Page 1: ...asurement instrument that can be used on products of virtually any size and in industries as diverse as paint and textiles It is available in two different models 45 0 and diffuse 8 sphere The label o...

Page 2: ...13 2 ColorFlex 07 02 The diffuse model...

Page 3: ...the sample port during standardization Green tile used to check instrument performance Use of the green tile is described on page 24 3 Standards care card gives instructions on how to clean the standa...

Page 4: ...nge juice tube holder 45 0 instruments only HL D02 1011 127 Citrus option 45 0 instruments only HL UNI Citrus CFLX OJ4 orange juice standard uncalibrated HL A11 1011 509 Set of 6 orange juice standard...

Page 5: ...he spring loaded clamp away from the sample port insert the sample and gently release the clamp Port Inserts 45 0 instruments only Port inserts with various sized openings may be purchased and placed...

Page 6: ...nd connector of the scanner wand into the AC power supply connector 4 Plug the AC power cord into an electrical outlet 5 Attach the wand holder to the ColorFlex at a convenient location Bar Code Reade...

Page 7: ...equence ENTER ROMAN NUMERAL IV A EXIT Note The information contained in this section is specific for use with the ColorFlex sensor For additional information regarding features maintenance or use of t...

Page 8: ...Cup Port Plate 45 0 instruments only This port insert is specially designed to accommodate the 2 5 inch 64 mm glass sample cup and to hold it in the proper position at the measurement port Remove the...

Page 9: ...ion is facing upward Insert the alignment pin on the tube holder into the small alignment hole at the top of the instrument port When the pin is properly installed into the hole the tube holder should...

Page 10: ...olor data displays Tomato Color Score Tomato Juice Tomato Paste Tomato Sauce Tomato Catsup and Tomato A B Ratio Install the indices after installation of Universal by copying the CMRINDEX CDX file con...

Page 11: ...nsert by pulling it out of the port insert retainer Replace it with the glass port insert Make sure that the flat portion of the insert is facing outward with the beveled portion toward the interior o...

Page 12: ...nter The external serial printer connects to the serial port of the ColorFlex and prints the setup data when the Read key is pressed and held or the measurement data when the Up Arrow Print key is pre...

Page 13: ...terLab HunterLab Part Number C04 1005 631 for adapter A13 1002 655 for 110V cord A13 1002 656 for 220V cord Be certain that the power cord is in good condition before connecting it The ColorFlex is gr...

Page 14: ...th 45 0 geometry requires reading of the black glass and white tile Standardization on a diffuse 8 ColorFlex requires you to read both the light trap and the white tile Standardizing a diffuse ColorFl...

Page 15: ...formation about your ColorFlex sensor and its standardization is given in its separate ColorFlex User s Guide and in the Sensor chapter of this manual It is recommended that the instrument be standard...

Page 16: ...ot the tiles dry using a clean non optically brightened lint free paper towel Allow the tiles to stabilize to room temperature before use Note SPARKLEEN is manufactured by Fisher Scientific Co Pittsbu...

Page 17: ...Guide for an explanation of the various possible error messages Resetting the ColorFlex On rare occasions your ColorFlex may display unusual characters or otherwise act strangely This is sometimes ca...

Page 18: ...rated under temperature conditions of 50 to 110 F 10 to 43 C Do not leave ColorFlex in any area where temperature extremes are possible Physical Characteristics Weight 11 25 lb 4 5 kg Dimensions 36 cm...

Page 19: ...character to ColorFlex Data flow begins again when ColorFlex receives an XON character Conditions of Illumination and Viewing Source Dual beam xenon flash lamp 500 000 flashes Source UV Content Nomina...

Page 20: ...nm 400 700 Photometric Range 0 150 Spectral Resolution 10 nm Spectral Bandpass 12 nm Measurement Speed at 23 C Warm up time 1 second Command to data output 1 second Read to read 3 seconds Warm up Per...

Page 21: ...rage DE value is less than or equal to 0 04 Contact HunterLab Customer Support as described on page 25 6 if the instrument fails the repeatability test Note Performance specifications are subject to c...

Page 22: ...55011 CISPR 11 1990 09 Class A Group 1 EN50081 1 1992 EN50082 1 1992 EN55011 1994 Class A IEC 801 2 1984 8 KV IEC 801 3 1984 3 V m 27 500 MHz IEC 801 4 1984 1 KV IEC 801 5 1991 2 KV Supplementary Info...

Page 23: ...nd textiles Because of its compact design it can be used to measure objects that would be difficult to position at the sample port of a larger color instrument Sample measurement methods can be found...

Page 24: ...13 24 ColorFlex 07 02...