Intended Use
The operator may only perform the maintenance work described in this manu-
al.
Use only the parts prescribed in this manual for servicing.
The tests and maintenance work defined by the manufacturer should be performed
to make sure that the operator remains safe and that the instrument continues to
function correctly.
Any service and maintenance work not described in this manual must be performed
by authorised service personnel and technicians.
Any changes on the system which are not authorised by the manufacturer lead to the
loss of guarantee obligation.
The system was developed and tested according to the regulations of the IvD
directive
Any changes on the system which are not authorised by the manufacturer lead to the
loss of the validity of the conformity with the applicable regulations the manufacturer
has declared. In this case, the customer is responsible for the fulfilment of the appli-
cable regulations.
The instrument should be opened, serviced and repaired by qualified and au-
thorised service personnel only.
1.2.2
Electrical Safety
Country-specific regulations regarding safe electrical operation of the in-
stalled system have to be observed.
Check that the operating voltage is set correctly on all system components be-
fore you connect the system to the mains supply.
This product must be operated from the type of power source indicated on the type
label. If you are not sure of the type of power available, consult an authorised sales
person or your local electric power company.
Use a 3-wire grounding type plug to connect all instrument and peripherals to
mains supply.
Use only extension cables with a protective conductor and grounded contact.
Grounding of the instrument, computer and its peripheral devices to the same pro-
tective earth potential must be ensured.
Never knowingly interrupt the grounding contacts.
There is the risk of an electrical shock if the protective conductor is interrupted within
or outside the device, or has been disconnected.
The instrument must be connected with a delivered connection cable to dedi-
cated socket. The use of a multi plug is not allowed!
Do not allow anything to rest on the power cord.
If you can see that the instrument has become unsafe to use, switch it off and
disconnect it from the mains supply.
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...