Emergency OFF

Pull Cord Switch

OVERHEAD LIFT

Power

ON/OFF

Emergency 

down

Handset

If lift does not operate and POWER 

LED does not light, pull cord once 

gently to turn power ON. May be 

used to cut power to lift entirely in 

an urgent situation.

UP/DOWN

Auxiliary POWER/

Emergency DOWN

switches

Auxiliary

UP/DOWN Switches

Manual Emergency

Lower

POWER - GREEN light indicates 

power ON. DIAGNOSTIC - Flashes 

RED in various sequences indicating 

various internal trouble situations. 

Solid RED light indicates low battery 

condition. Battery should be charged 

immediately. CHARGING - YELLOW 

indicates active charging. GREEN 

indicates fully charged.

For use if Handset has

malfunctioned.

EMERGENCY DOWN

may be used to lower

patient in event of

malfunction in lift

circuitry.

May be used in event handset

does not operate.

May be used in event of power 

failure or lift malfunction.

HELPFUL HINT

Return to charge after every use.

Note: some hand control models will 

have RTC (return to charge) button. 

Motor will attempt to return to docking 

station and if track configuration is a 

H-track you will need to manually push 

motor to charging dock.

READY TRACK REFERENCE GUIDE