OWNERS MANUAL | Cobolt Modulated DPSS Lasers | D0342-C JUNE 2016
Communication commands
The laser is delivered with the Controller set in Auto-start mode (see section 5.1 for Auto-start sequence description).
For system integration the Auto-start sequence can be disabled and the following commands can be used to control
the laser (NOTE some commands require Auto-start to be disabled but others will work when Auto-start is active). As
long as power is supplied to the Controller the temperature control elements are always operating to reach set-point
values and the laser will be idle waiting for the next command. All arguments are in lower case and separated by a
space (ASCII 32).
Command Function
Argument
Returned value
ilk?
0 = OK, 1 = remote
interlock open
@cob1
Laser ON after interlock Forces the laser
into Autostart without checking if
autostart is enabled (OEM models).
@cobas
Enable/disable autostart
See sect 5.1 for description (to be used
on OEM models only, disabling
autostart means the laser no longer
goes through its warm up routine).
0 = disable, 1 = enable
@cobas?
Get autostart enable state
0 = disabled, 1 = enabled
l?
Get laser ON/OFF state
0 = OFF, 1 = ON
l1
Laser ON
Requires autostart disabled. Use this
command for manual ON (OEM
models).
l0
Laser OFF
Use this command for manual OFF
(OEM models).
p?
Get set output power
Float (W)
p
Set output power
Float (W)
(e.g. p 0.050 for 50 mW)
pa?
Read output power
Float (W)
i?
Get drive current
Float (A)
slc
Set drive current
Float (A)
leds?
Status of 4 LEDs
Int [0:15]
Bit 0 = “POWER ON”
Bit 1 = “LASER ON”
Bit 2 = “LASER LOCK”
Bit 3 = “ERROR”
1 = LED on
0 = LED off
f?
0 = no fault
1 = temperature error
3 = open interlock