11 | 60
OWNERS MANUAL | Cobolt 08-01 Series | D0141-K SEPTEMBER 2020
7.
The laser will now start up in continuous-wave, constant current mode at its nominal maximum power level.
The power and wavelength may continue to drift for up to 3 minutes while the temperature of the platform
stabilizes.
NOTICE
If the power does not match the power as stated on the test sheet see Section 12: Service
for more information.
06-DPL
When delivered the lasers are, by default, set to continuous-wave, constant power mode. As soon as power is supplied
to the laser head the auto-start procedure will begin. Light will be emitted once the remote interlock connector is
connected, the shutter is open and when the key is turned to the ON position (CDRH model).
1.
Mount the laser on a heat sink or suitable flat surface that provides adequate heat dissipation and connection
to ground. Use the four holes on the laser’s base plate to secure it.
2.
Attach the 6-pin Molex cable to the laser head. Be sure the orange arrow is facing the top (labelled) side of
the laser head.
3.
Attach the 15-pin D-SUB end of the cable to the key control box.
4.
Insert the remote interlock plug into the connector on the key control box.
5.
Connect the supplied 5 V power supply unit to the socket on the laser head and plug it in to the mains.
6.
To start the laser, turn the key on the control box clockwise to the ON position. If it is already in the ON
position, turn it to OFF and then ON again.