Commissioning
Chapter 2
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
31
Tighten the lock screws on the leveling feet after aligning the temperature control unit. This
prevents the leveling feet from changing their height during operation.
2.10.2
Opening/closing valves
Opening valves:
Open valves by turning them counterclockwise (turn 90° left as far as it will go).
Close valves:
close valves by turning them clockwise (turn 90° right as far as it will
go).
PROCEDURE
Check that all valves are closed.
Close all valves by turning them clockwise (turn 90° right as far as it will go).
2.10.3
Emergency stop switch (optional): Check switching status
PROCEDURE
Check the
>Emergency stop switch<
[70].
This switch must not be pressed (activated) when taking the temperature control unit into oper-
ation.
If required, release the
>Emergency stop switch<
[70] by turning it clockwise. A built-in spring in
the
>Emergency stop switch<
[70] returns it to its original state.
2.11
Connecting externally closed application
The illustration “connection diagram” can be found on page 85 in section
.
2.11.1
Connecting an externally closed application
Overpressure in the application (e.g. > 0.5 bar (g) with glass apparatus)
MATERIAL DAMAGE TO THE APPLICATION
Provide an overpressure protective device to prevent damage to the application.
Do not install valves/quick-release couplings in the feed/discharge lines from the temperature
control unit to the application and from the application to the temperature control unit.
If valves/quick-release couplings are required:
Install burst disks on the application itself (at the feed and discharge lines).
Install a bypass upstream of the valves/quick-release couplings for the application.
Matching accessories (e.g. bypasses to reduce pressure) can be found in the Huber catalog.
Opening and closing
valves
Exemplary layout of
the switches