Commissioning
Chapter 2
OPERATION MANUAL
V2.3.0en/07.06.19//17.12
Unichiller®
23
2.1.1.2
Temperature control unit without lifting eyes
▪
Do not lift and transport the temperature control unit alone and without aids.
▪
Lift and transport the temperature control unit only with an industrial truck.
▪
The industrial truck must have a lifting force equal to or greater than the weight of the tempera-
ture control unit. See the data sheet (from page 85 in section
) for the weight of the
temperature control unit.
2.1.2
Positioning the temperature control unit
2.1.2.1
Temperature control unit with casters
▪
Do
not
use the casters for transportation to the place of installation. Observe page 22, section
»Lifting and transporting the temperature control unit«
for the transport to the place of installa-
tion.
▪
Use the rollers only for positioning at the place of installation.
▪
Only ever move the temperature control unit on casters if the surface is level, without a gradient,
non-slip and stable.
▪
Do not move the temperature control unit alone.
▪
At least 2 persons
are required to move the temperature control unit on casters.
At least 5 per-
sons
are required to move the temperature control unit on casters if the total weight of the tem-
perature control unit is
over 1.5 tons
.
▪
The parking brakes must be activated at the casters and/or the leveling feet (if any) must be un-
screwed/activated before the temperature control unit is put into operation (see page 30, section
»Unscrewing/activating the leveling feet (if any)«
).
2.1.2.2
Temperature control unit without casters
▪
An industrial truck must be used for positioning the temperature control unit.
▪
Do not move the temperature control unit alone.
▪
At least 2 persons
are required to move the temperature control unit.
▪
The industrial truck must have a lifting force equal to or greater than the weight of the tempera-
ture control unit. See the data sheet (from page 85 in section
) for the weight of the
temperature control unit.
▪
The leveling feet (if any) must be unscrewed/activated before the temperature control unit is put
into operation (see page 30, section
»Unscrewing/activating the leveling feet (if any)«
).
2.2
Transportation lock
Commissioning with active transportation lock
DAMAGE TO THE TEMPERATURE CONTROL UNIT
Check the position of the transportation locks.
Prior to commissioning of the temperature control unit, bring the transportation locks of the
compressor into the operating position.
The temperature control units are listed in the table below are equipped with transportation locks
for the compressor. These transportation locks must be unlocked to commissioning the temperature
control unit or be reactivated during transport to another place of installation.
Special or temperature control unit variants may vary from those listed in the table. These tempera-
ture control units have a sticker next to the nameplate that indicates the type of transportation lock
used.
Temperature control units
Transportation lock
- All Unichiller table models
- Unichiller series: 0xxT(w)/(w) EO; 1xxT(w)/(w) EO; 2xxT(w)/(w) EO
without
- Unichiller series: 3xxT(w)/(w) EO; 4xxT(w)/(w) EO; 5xxT(w)/(w) EO
Type A
Overview of transpor-
tation locks