Introduction
OPERATION MANUAL
Chapter 1
RotaCool®
V2.3.0en/04.08.21//1.30
18
1.5
General information
1.5.1
Description of workstation
The workstation is located at the control panel in front of the temperature control unit. The work-
station is determined by the customer's connected peripheries. Accordingly, it must be designed
safe by the responsible body. The workstation design also depends on the applicable requirements
of the German occupational health and safety regulations [BetrSichV] and the risk analysis for the
workstation.
1.5.2
Safety devices to DIN 12876
The rating of your temperature control unit is stated on the data sheet in the appendix.
Classification
Temperature
control medium
Technical requirements
Identification
d)
I
Non-combustible
a)
Overheat protection
c)
NFL
II
Combustible
b)
Adjustable overheat protection
FL
III
Combustible
b)
Adjustable overtemperature protection
and additional low-level protection
FL
a)
Usually water; other fluids only if non-combustible even within the temperature range of an individual fault.
b)
The temperature control media must have a fire point of ≥ 65 °C.
c)
The overheat protection can, for instance, be realized using a suitable fill level sensor or a suitable temperature
limiter.
d)
Optional at the choice of the manufacturer.
▪
Temperature control units with heating correspond to class number III/FL. These temperature
control units are characterized by an “H” in the device name.
▪
Temperature control units without heating correspond to class number I/NFL.
Rating of laboratory
thermostats and
laboratory baths
Overview of the tem-
perature thresholds