background image

3

● Current test range

10mA

● current measure precision

5%+0.2nA

● short-circuit current

5mA

● Capacitance test range

20uF

● Capacitance testing accuracy

15%+0.03uF

● Insulation resistance testing range and accuracy

T:23±5ºC

RH :45 – 75%

range

accuracy

500V

1000V

2500V

5000V

10000V

Unspecified

<1M

<1M

<1M

<1M

<1M

5%

1M-10G

1M-20G

1M-100G 1M-100G

1M-200G

20%

10G -100G 20G-200G

100G-1T

100G-1T

200G-2T

Unspecified

> 100G

> 200G

> 1T

> 1T

> 2T

range

accuracy

15000V

Unspecified

<1M

5%

1M -300G

20%

300G -3T

Unspecified

> 3T

IV. Introduction of Instrument

The Front Panel Description

Summary of Contents for HZJY-15K

Page 1: ...HZJY 15K Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...Contents I Warning 1 II Feature 2 III Specification Parameter 2 IV Introduction of Instrument 3 V Instrument Using 7 VI Packing List 8...

Page 3: ...erators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in ele...

Page 4: ...eaches 2 mA still ensure accuracy of test instruments 2 methods for power use of lithium ion batteries battery life can be up to 4 hours 10000V 100M Insulation resistance simulation column Digital fil...

Page 5: ...tance testing range and accuracy T 23 5 C RH 45 75 range accuracy 500V 1000V 2500V 5000V 10000V Unspecified 1M 1M 1M 1M 1M 5 1M 10G 1M 20G 1M 100G 1M 100G 1M 200G 20 10G 100G 20G 200G 100G 1T 100G 1T...

Page 6: ...iption OFF Power Off No operation 3 minutes automatic shutdown H V Set test voltage Manual 500V 1000V 2500V 5000V 10000V 15000V TIME SetThe test time 01 00 10 00 30 00 Switching software keys to a new...

Page 7: ...p 1 In shutdown state hold key 2 the 2 key 2 Press the 1 key the power button until into the contrast adjustment menu 3 Press 1 or 2 keys to adjust contrast until the right 4 Press the 4 key save and...

Page 8: ...Regulating key Soft key Description Regulating key By a voltage rise of 10V long time boost up the fastest up to 600V per second Regulating key According to a voltage drop 10V long time step down spe...

Page 9: ...e polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connection points for attention 1 Tested to confirm product safety grounding try not charged 2 Confirm the...

Page 10: ...8 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Red test line 1 4 Black test line 1 5 Green test line 1...

Reviews: