background image

5

2.3.2 Adjustment

2.3.2.1

Pull the screw axle by hand to observe the flexibility of the sliding unit. If any trouble, inspect

at each transmission joints for any loosing or barrier. Contact us if the trouble cannot be solved.

2.3.2.2

Move the stirrer and observe if it can touch the thermometer, if yes, adjust the position of

blade.

2.3.2.3

Gas source connection: Turn off the master gas valve and adjusting valves, connect the gas

source with the master gas valve. When igniting, turn on the master valve (Fully or half open

depending on the gas pressure), and turn left the igniting adjusting valve slowly to adjust diameter

of flame to standard size.

Note: When igniting, dispose the air in the tube and then adjust the valve slowly, carefully

control the size of flame within stated range, so as to avoid fire hazard.

III.Operation

3.1 Commissioning

3.1.1 Power supply

Turn on “Power” switch and the light works.

Liquid return joint and
pipe

Gas inlet

In/Out liquid joint with
valve

Summary of Contents for HZBS-002

Page 1: ...HZBS 002 Abel Flash Point Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nd operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrume...

Page 3: ...Contents I General 1 II Installation and commissioning 2 III Operation 5 VI Trouble and trouble shooting 7 V Maintenance 8...

Page 4: ...d by SSR which can satisfy requirements of different heating rate Continuously stirring the sample and stirring is auto stopped during igniting Electrical controlled ignition Test cup lid with stirrin...

Page 5: ...mperature rise rate Stirring rate can be adjusted in range of 0 40 roll min by stirrer adjust button Stirrer will be stop working during ignition 1 3 3 Gas source Beside master gas valve we have adjus...

Page 6: ...2 2 1 The instrument should be placed on steady test platform with no airflow around 2 2 2 The distance between instrument and the walls or other shields should be not less than 200mm 2 3 Installation...

Page 7: ...t gas source into gas inlet and fix well If test in low temperature connect intake pipe to In Out liquid joint connect liquid return pipe to overflow pipe and fix well If test not in low temperature y...

Page 8: ...alve and adjusting valves connect the gas source with the master gas valve When igniting turn on the master valve Fully or half open depending on the gas pressure and turn left the igniting adjusting...

Page 9: ...point of sample around 15 user can use icy water as medium and if flash point around 40 water with normal temperature can be used Switch off the water inlet cover and funnel the medium into the bath S...

Page 10: ...will be descended 3 2 4 When test operation is complete adjust voltmeter back to 0 turn off RUN switch off gas source remove all liquid medium in the bath VI Trouble and trouble shooting Trouble Caus...

Page 11: ...haft and stationary shaft homocentric Fire go out when ignition 1 Test flame too small 2 Disturbing air flow 1 Adjust flame size 2 Prevent air flow V Maintenance 5 1 The instrument should be placed at...

Reviews: