resistance of the closing coil to confirm that the closing coil current is large. Then please check the
wiring carefully, make sure that the closing output is not short-circuited, and then restart after power off.
9.3 The instrument has no DC power output
This kind of situation may be caused by the burnout of the power supply insurance of the instrument
and the loss of power supply to the entire instrument. At this time, the replacement insurance can be
replaced. Or the internal power supply of the instrument has failed. At this time, external triggering can be
used for measurement. After the test is completed, notify the company to return to the factory for repair.