25
XI.Specifications
All specifications apply from +18°C to +28°C unless stated otherwise. All specifications
assume a 5-minute warm-up period. Temperature coefficient from -10°C to +18°C and
+28°C to 55°C is ± 0.005%/°C of range per °C.
All specifications are not based on clearing except resistance and RTDs. To achieve
higher accuracy, please perform clear function to eliminate offset errors.
Maximum Input Voltage
:
30V.
DC Voltage
Range
Accuracy
Measurement
(0~100)mVDC(Upper display)
±0.02%
(0~30)VDC(Upper display)
±0.02%
(0~100)mVDC(Lower display)
±0.02%
(0~20)VDC(Lower display)
±0.02%
Source
(0~100)mVDC
±0.02%
(0~10)VDC
±0.02%
Resistance
Range
Accuracy
4-wire
2-, 3-wire
Accuracy
Accuracy
Measurement
(0~400)Ω
±0.1Ω
±0.15Ω
(0.4~1.5)kΩ
±0.5Ω
±1.0Ω
(1.5~3.2)kΩ
±1.0Ω
±1.5Ω
Excitation Current
:
0.5mA
Clear of resistance before measuring according to ‘8.4 Clear of Resistance and
RTDs’.