23
(1)
Press right
to select the decided TC.
(2)
Short the test leads with Measure jacks.
(3)
Press right
after the value is stable.
8.4 Clear of Resistance and RTDs
To clear offset associated with the test lead resistance in 2-wire, 3-wire or 4-wire ohms
measurements, proceed as follows:
To clear offset of measuring or sourcing resistance, proceed as follows:
(1)
Press
to select Measure or Source mode.
(2)
Press
to select the decided resistance.
(3)
Short the test leads with Measure or Source jacks.
(4)
Press right
after the value is stable
To clear offset of measuring or sourcing RTD, proceed as follows:
(1)
Press
c to select Measure or Source mode.
(2)
Press
to select the decided RTD.
(3)
Short the test leads with Measure or Source jacks.
(4)
Press right
after the value is stable.
8.5 Clear of Sourcing TC
To clear offset of sourcing TC must be according to ‘10.2 Clear of Sourcing Voltage’.
IX.Calibrating a Transmitter
Use Measure mode (upper display) and Source mode (lower display) to calibrate the
transmitter.
Proceed as follows to take calibration of a transmitter.
(1)
Press
until the upper display shows “MEASURE LOOP”.
Summary of Contents for EDEKA TC-200