17
process calibrator.
Proceed as follows to simulate transmitter:
(1)
Press
until the display shows “SOURCE”.
(2)
Press
to select simulating transmitter until the display shows “LOOP mA”. The
EDEKA TC-200 Multi-function process calibrator produces a variable resistance to
regulate current to the specified value.
(3)
Connect an external loop power supply as shown in Figure6- 7.
(4)
Enter the desired output value by pressing numeric keypads and press
.
Figure6- 7 Connections for Simulating a Transmitter
V. Cold Junction
Two modes are available for setting cold junction.
1) Automatic Mode
When you turn on the EDEKA TC-200 Multi-function process calibrator, it is just in the
automatic mode. At this mode, the EDEKA TC-200 Multi-function process calibrator takes
the temperature of jacks as cold junction temperature.
2) Manual Mode
Summary of Contents for EDEKA TC-200