11
The hipot tester comes with power safety detection function, there is an "OK" indicator
lamp on the front panel, if the "OK" indicator both lights, power input is correct, you can
press the start button, if either “o” or “k” not bright, the input power is error, please do not
start the hipot tester and check the input power supply.
(
2
)
Timer
When the test time is set (greater than 0), after the test is begun, the counter will begin to
reverse, after the time value reached 0,the tester will cut off the output voltage
automatically. When the test time is set to "0.0s", the tester will continue testing until user
presses the "STOP" button or set the “reset” signal enable. Anyway, whether it is time
testing or continuous testing, users can stop the tester by pressing the "STOP" button or
enable the “reset” signal .
(
3
)
Remote interface
、
PLC interface
This hipot tester can be equipped with remote control gun or the PLC interface to start or
reset.
(
4
)
Danger indicator
To ensure operator safety, the danger indicator will be lit if there is high voltage. After the
hipot tester starts to test, If the output voltage is lower than the setting voltage, the danger
indicator will flash. If the output voltage is higher than the set voltage, the danger indicator
will bright normally. Even the withstanding voltage tester is in the stop status, if high
voltage output is larger than the setting voltage, the danger indicator will light to warn
there is high voltage output, users should not contact any object in the H.V output ports to
prevent electric shock.