WASHER-DISINFECTOR 

CLEANING MONITOR

For Use in Routine Monitoring of the Wash Phase 
in Washer/Disinfectors

IMS-1200W

  |  

50 TESTS

INDICATIONS FOR USE: 

The single use Hu-Friedy Washer-Disinfector Cleaning Monitor is designed for monitoring cleaning efficiency during the wash process.

INSTRUCTIONS FOR USE: 

1.   Insert a Washer-Disinfector Cleaning Monitor strip (IMS-1200W) into Cleaning Monitor Holder (IMS-1200H) completely. The front side of the monitor with 
  red test soil will be visible in the holder window, as shown below. 

2.   Position the Cleaning Monitor Holder face up inside a basket. Place basket on a shelf in the washer. Vary positions and shelves of Washer-Disinfector 

  Cleaning Monitor placement cycle to cycle. 
3.   Run a complete washer cycle. We recommend monitoring every shelf on every washer each day for a machine release and using a minimum of 

  one Washer-Disinfector Cleaning Monitor strip per load for routine monitoring throughout the day. 
4.   At the end of the washing cycle, remove the Washer-Disinfector Cleaning Monitor from the holder and read the result. Please see visual interpretation of 
  results below. Caution: Holder and monitor strip may still be hot! Remove with care. 

INTERPRETATION OF MONITOR RESULTS:

5.   If desired, you may also record the location of monitor placement in the washer-disinfector on the strip itself with a fine tip waterproof marker. 

  Place a hash mark or circle representing the position of the Wash-Checks on the location rectangle that simulates an imaginary washer basket. 

  This allows for placement tracking. 
6.  If desired, staple the Washer-Disinfector Cleaning Monitor strip onto the Washer-Disinfector Cleaning Monitor Record Sheet and complete the 
  appropriate information. Retain as a permanent record. 

STORAGE: 

•   Store Cleaning Monitors inside the protective foil bag provided at 40-86°F (4-30°C). Do not remove the monitors from the foil bag until ready to use.
•  Keep away from disinfectants and fumes. 

EXPIRY DATE:

The expiry date is printed on the product packaging.

Please see website for Washer-Disinfector Cleaning Monitor Record Sheet and Troubleshooting Guide.

1-800-Hu-Friedy

Hu-Friedy.com

Manufactured for:
Hu-Friedy Mfg. Co., LLC
3232 N. Rockwell St. 
Chicago, IL 60618 USA

Hu-Friedy Mfg. Co., LLC
Zweigniederlassung Deutschland
Kleines Öschle 8
D-78532 Tuttlingen

Untreated Monitor, or critical/poor result 

Insufficient result 

Optimum result 