COMBITEST422
EN - 58
6.7.4. Br.Cap mode
– Verification of the breaking capacity of protection devices
1. Press the
MENU
key, move the cursor to
LOOP
in the
main menu by means of the arrow keys (
,
) and
confirm with
ENTER
. Subsequently, the instrument
displays a screen similar to the one reported here to the
side.
Select
“Europe” as a country (see § 5.1.2), the options
“TN, TN or IT”, “25 or 50V”, “50Hz or 60Hz” and the
reference voltage in the general settings of the instrument
(see § 5.1.3).
L O O P
15/10
– 18:04
TN
max
psc
I
ZL-L
=
=
- - -
- - -
A
FREQ. = 50.00Hz
VL-PE=0V
VL-L=0V
L-L
Br.Cap
15kA
FUNC
MODE
Lim
2. Use the
,
keys to select the parameter to be modified, and the
,
keys to modify the
parameter value.
➢
FUNC
→
the virtual key allows setting the measuring mode of the instrument, which
may be:
L-N, L-L
or
L-PE
.
➢
MODE
→
the virtual key
allows setting the instrument’s operating mode. Select the
Br.Cap
option.
➢
Lim
→
the virtual key allows setting the maximum tripping current, expressed in
"kA", at which the protection must break the circuit, in the range:
0.1kA ÷ 999kA
.
3. If possible, disconnect all loads connected downstream of the point to be measured, as
the impedance of these users could distort the test results. Perform the preliminary
calibration of the test leads as described in § 6.7.2
4. Insert the green, blue and black connectors of the three-pin plug cable into the
corresponding inputs B3, B4 and B1 of the instrument. As an alternative, use the single
cables and apply the relevant alligator clips to the free ends of the cables. It is also
possible to use the remote switch probe by inserting its multipolar connector into the
input B1. Connect the Mains Plug, the alligator clips or the remote switch probe to the
electrical mains according to Fig. 22, Fig. 23, Fig. 24, Fig. 25 or Fig. 26.
5. Note the presence of the correct voltage values between
L-L and L-PE corresponding to the selections carried out
in the initial phase (see § 5.1.3) as shown in the screen to
the side.
L O O P
15/10
– 18:04
TN
max
psc
I
ZL-L
=
=
- - -
- - -
A
FREQ. = 50.00Hz
VL-PE=223V
VL-L=387V
L-L
Br.Cap
15kA
FUNC
MODE
Lim