GSC53N - GSC57 - ZG47
EN - 13
5.4.
RESET
This option re-establishes the initial settings of the instrument in
ANALYSER
function.
The RESET command re-establishes the limit value on Insulation measurement (see
paragraph 6.2) at 0.5M and test voltage at 500V.
The “not modified” parameter it is not modified by RESET command
The initial settings of the instrument consist of:
ANALYSER CONFIG:
System:
3PH4W
Frequency:
not modified
Current range:
not modified
Clamp type:
not modified
Transforming ratio of voltmetric transformers:
1
Password:
OFF
RECORDER CONFIG:
Start:
MANU (the recording is started
at 00 sec mark on clock after pressing
the
START/STOP
key)
Stop:
MANU
Integration period:
15min
Recording of harmonics:
ON
Recording of Voltage anomalies (Sag and Surge:
ON
Voltage Reference for Sag and Surge detection:
230V
Upper Limit for Sag and Surge detection:
6%
Lower Limit for Sag and Surge detection:
10%
Selected voltages:
V1, V2, V3
Selected voltage harmonics:
THD, 01, 03, 05, 07
Selected currents:
I1, I2, I3, IN
Selected current harmonics:
THD, 01, 03, 05, 07
CO-GENERATION:
OFF
Powers, Pf and cos selected:
Pt, P1, P2, P3
Qti, Q1i, Q2i, Q3i
Qtc, Q1c, Q2c, Q3c
St, S1, S2, S3
Pft, Pf1, Pf2, Pf3
dpft, dpf1, dpf2, dpf3
Energies:
Eat, Ea1, Ea2, Ea3
Erit, Eri1, Eri2, Eri3
Erct, Erc1, Erc2, Erc3
The RESET command will not
erase the instrument’s memory.
Summary of Contents for GSC53N
Page 1: ...Copyright HT ITALIA 2012 Release EN 1 08 28 02 2012 GSC53N GSC57 ZG47 Instruction Manual...
Page 2: ......
Page 151: ......