___________________________________________
22
ENGLISH
EDIT
The “free programs” section can be accessed during the “program
selection” phase by pressing the EDIT key. The relative LED lights up.
Through EDIT you can customize up to 10 programs.
The editable parameters for each program are:
• DURATION - from 1 to 59 minutes
• FREQUENCY - 1 MHz or 3 MHZ
• POWER - from 5% to 100% of the power
• DUTY CYCLE - from 5% to 100% of the Duty Cycle
During the parameter setting phase, the display will show the name of the
free program being edited and each time the variable on which you are
working:
W
ith the “▲” and “▼ ” buttons you can select the variable (eg duration),
w
ith the “
+
” and “-” buttons you can set the value (eg 09 minutes).
Pressing the EDIT
k
ey again, the data of the free program, if modified, are
saved. A
"
beep
"
w
ill confirm the saving and the EDIT LED
w
ill turn off.
During the EDIT phase the START / STOP
k
ey is not active.
F R E E
1
D U R A T I
= 0 9 M I N
F R E Q U E N C Y = 1 M H Z
1
P O W E R
= 0 5 %
1
D U T Y
C Y C L E = 0 5 %
F
EEE
1
EE
R
F R E E
F R E E
O N
!
Caution
APPLICATION METHOD
The application with the appliance is simple, but in order to obtain satisfactory results
it is advisable to perform the sessions taking into account some fundamental rules.
We remind you that the air attenuates the ultrasounds, so it is always necessary to
interpose a certain quantity of ultrasound gel or cream suitable for massage,
between the treatment head and the surface of the skin to be treated. The treatment
head must be applied perpendicular to the area to be treated and slid slowly, without
too much pressure, collecting the gel or cream used from time to time always
keeping a certain layer of gel or cream between the treatment head and the skin . To
avoid overheating the treatment area it is necessary to move the head with a slow
and uniform movement, taking care never to leave it stationary.
- If it is necessary to make an "unexpected" interruption, remember to
"ALWAYS" pause the machine, otherwise the head LEFT RUNNING, immobile
and without creams, will overheat with the risk of damage;
- before proceeding with the application it is advisable to identify the part of the
body on which to act. For anti-cellulite treatments we recommend the upright
position for a correct evaluation of the area to be treated, because in a sitting or
lying position there is a staggering of the silhouette;
- having identified the exact area, select the type of program and the action on
the device;
- the actions that the program proposes foresee application times on small
areas, therefore to act on larger areas it will take a longer amount of time,
having to repeat the same program several times depending on abstention of
the area to be treated;
FUNDAMENTAL RULES TO RESPECT
!
Caution
Handle the treatment head with care. Careless handling can affect its
characteristics.
Do not knot or twist the connection cable.
Do not use in the vicinity of the device and the treatment head, sprays,
liquids or solvents.
- before starting the program, pour a large quantity of gel or a specific cream
on the treatment head;
- start the program and place the treatment head on the bare part to be treated;
- move the treatment head in a slow circular motion, back and forth
or as for drawing an eight. Depending on the program, you will feel a slight
vibration or a slight heat.
END OF TREATMENT
At the end of the treatment:
- turn off the appliance by turning the main switch to "0" (zero). If the
appliance remains on, its operation is not compromised;
- disconnect the treatment head from the device panel, taking the connector
and pulling upwards perpendicularly, avoiding tearing;
- disconnect the connector plug from the socket;
- disconnect the power cable from the back of the device by pulling it
outwards with very slight vertical oscillations;
- clean the treatment head, avoiding the use of solvents or abrasives, as
indicated in the "Maintenance" chapter;
- collect the treatment head and power supply cables properly
neat, avoiding knotting or twisting them;
- reposition the device and its accessories in the supplied case;
- store in a dry and safe place, out of the reach of children, avoiding
accidental falls.
!
Caution
At the end of each treatment, carry out the maintenance operations
described in the "Maintenance" chapter indicated with periodicity "A".
Maintenance
MAINTENANCE OPERATION
PERIODICITY
Check the integrity of the power cable.
M
F
Check the integrity of the treatment head cable.
M
F
Verification of the integrity of the treatment head.
A
A
Cleaning the appliance.
A.N.
A.N.
Cleaning and disinfection of the treatment head.
A
A
Periodic calibration
2Y
2Y
Electrical checks of the appliance.
2Y
2Y
REGULAR USE
FREQUENT OR INTENSE USE
For maintenance, only original products and spare parts must be used.
Improper use of non-original products and spare parts causes the
warranty to be void.
The information in this chapter is provided in order to allow the user to
obtain the best performance from the appliance in the best safety
conditions.
- never carry out any operation with damp or wet hands;
- do not immerse the appliance in water for cleaning, use a cloth soaked
in non-abrasive neutral detergent;
- do not try to open or repair the appliance. Contact the Authorized
Service Center for any type of intervention or check.
!
Caution
During maintenance of the appliance:
i
Information
Periodicity:
!
Caution
The biennial maintenance of the appliance must be carried out
exclusively by HSD S.r.l.
At each treatment.
A
Monthly.
M
Fortnightly.
F
A.N. As needed.
Biennial.
2Y
i
Information