CLEANING AND DISINFECTION
During maintenance of
the appliance:
- never carry out any operation
with damp or wet hands;
- do not immerse the appliance in
water for cleaning, use a cloth
soaked in non-abrasive neutral
detergent;
- do not try to open or repair the
appliance. Contact the Authorized
Service Center for any type of
intervention or control.
o
CAUTION!
It is forbidden to use chemical or
abrasives for disinfection. During
disinfection, avoid contact
between the connector and the
antiseptic solution used. Follow
the directions and precautions for
use provided by the manufacturer
of the antiseptic solution.
•
MAINTENANCE
CAUTION!
For any orders of accessories
spare parts, consult the
accessories catalog, requesting it
by e-mail or by consulting the
manufacturer's website.
CAUTION
!
Identification plates are affixed
to the rear of the device and on
the speakers with the technical
data to be communicated to the
Authorized Service Center, the
Manufacturer or the Dealer in
case of request for accessories
or spare parts.
CAUTION!
Replacement diffusers can only
be obtained from the selling
company. Using a diffuser other
than the one indicated can be
dangerous.
CAUTION!
The biennial maintenance of
the appliance must be carried
out exclusively by HSD S.r.l.
CAUTION!
Before each action, set the ignition
switch to "0" (zero) and disconnect
the appliance from the power outlet.
Handle the treatment diffuser with
care. Careless handling can affect
its characteristics.
Cleaning the appliance
Remove the dust on the outside of
the appliance using a soft cloth
moistened with water, or with an
aqueous solution consisting of 75%
water and 25% denatured ethyl
alcohol.
CAUTION!
It is forbidden to use abrasives or
chemicals for cleaning.
Cleaning the diffuser
of treatment
Disconnect the treatment diffusers
from the device and remove any
residual cream or gel using a soft
cloth moistened with water, or with
an aqueous solution consisting of
75% water and 25% denatured
ethyl alcohol. After cleaning, dry
well and remove any residual
moisture from the treatment
diffusers before storage or
subsequent use.
CAUTION!
It is forbidden to use abrasives or
chemicals for cleaning. When
cleaning, avoid contact between
the connector and the cleaning
solution used.
Disinfection of
treatment diffusers
To disinfect the treatment
diffusers, use a cotton ball or
gauze, lightly soaked in a non-
irritating and alcoholic antiseptic
solution (for example the type used
for small abrasions or grazes,
available in pharmacies).
Control of
treatment diffusers
At each cleaning and - or
disinfection of the treatment
diffusers and in any case at the
end of each treatment it is
necessary to check that there are
no cracks or fissures that could
allow the entry of conductive liquid.
If the treatment diffusers show
cracks or cracks, replace them
immediately.